PEAK PERFORMER 1 TCD: 940-105
Detection: Nitrogen, Argon and Oxygen
Matrix Gas: He, Ar, O2, H2 and N2
High sensitivity GC measurements & amp detection.
The Peak Performer 1 TCD uses a sensitive single filament Thermal Conductivity Detector (TCD) which has been designed specifically for process gas or laboratory applications.
The TCD is used to measure impurities or mixtures from ppm to percent levels. The PP1 TCD uses Peak Laboratories proven operating platform for simple, accurate measurements. The PP1 technology has proven to offer a more cost-effective and user friendly alternative to other process GC's. In addition, Peak Labs practical experience and ability to customize to suit your applications makes Peak Labs your partner, not just your supplier.
Please contact Peak Laboratories here or contact a local sales representative for custom applications and further information on our standard applications.
Technical information

The Peak Performer 1 TCD gas chromatograph (GC) can be optimized for your analytical needs in a variety of matrix gases. Typical applications for model 940-105 are provided below:
- Detection of Argon, Nitrogen and Oxygen compounds
(see Fig. 1). - Semiconductor Plants
- Quality Assurance / Control
- Process Control
- Purifier Manufactures
Performance*
Some standard detection limits (in parts per million)
Matrix Gas |
He |
Ar |
O2 |
H2 |
N2 |
|
---|---|---|---|---|---|---|
Impurity | ||||||
N2: Nitrogen | 100 | 100 | 100 | 100 | N/A | |
O2: Oxygen | 100 | 100 | N/A | 100 | 100 | |
Ar: Argon | 100 | N/A | 100 | 100 | 100 |
All performance specifications are based on fully optimized PP1 with 2.0 cc sample loop.
Accuracy
- + / - detection limit or 10% of reading, whichever is higher.
Range
- 10000 : 1
Operation
Run time:
- 300 seconds < (depending on application)
Operating Temperature:
- 55 - 85°F (13 - 30°C )
Typical Gas Supply Requirements:
- Carrier:
- Purified research grade helium
- Supply pressure 70 – 110 psig with 5% stability to purifier
- Total flow ~ 200 sccm (plus ~ 50 cc per actuation)
-
Actuator
- N2/Air or inert gas regulated @ 70 psig
Data Collection / Communication (Standard)
- Backlit 3" x 5.25" touch screen
- Visual chromatogram and numerical results
- On-board integration with re-run capability
- Peak Viewer software: chromatograms collected and archived via standard 9 – pin com-port in .csv format (data is excel compatible)
Dimensions / Electrical
- 27" L x 17" W x 7" H
- 25 lbs.
- 115 VAC, 50 – 60 Hz / 220 VAC, 50-60 Hz
- 1.5 amp maximum
Accessories and other features
- Analog (VDC or mA) trend outputs
- Analysis averaging
- Auto start/stop capability
- 4-20 mA Output
- On Column Syringe Injector Adapter
- Dual Sample Stream
Figure 1.
Carrier: Purified Helium
Sample : Nitrogen, Argon and Oxygen within He, Ar, O2, H2 and N2 matrix gases
Please contact Peak Laboratories here or contact a local sales representative for custom applications and further information on our standard applications.