Peak Performer Gas Analysis


PEAK PERFORMER 1 RCP: 910-135

Reducing Compound Photometer

Detection: Hydrogen and Carbon Monoxide

Matrix Gas: Ammonia

 

The Peak Performer 1 uses a reducing compound photometer (RCP) to measure trace impurities down to part per billion levels, while offering a wide linear range for simple, accurate measurements.


Peak Labs Data Sheet Performer TCD

The Peak Performer 1 uses a reducing compound photometer (RCP) to measure trace impurities down to part per billion levels, while offering a wide linear range for simple, accurate measurements.

The PP1 technology has proven to offer a more cost-effective and user friendly alternative to other instruments for similar applications. In addition, Peak Labs practical experience and ability to customize to suit your applications makes Peak Labs your partner, not just your supplier.

Please contact Peak Laboratories here or contact a local sales representative for custom applications and further information on our standard applications.


Contact Us Online


Technical information

technical information tab

The Peak Performer RCP gas chromatograph (GC) can be optimized for your analytical needs in a variety of matrix gases. Typical applications for model 910-135 are provided below:


  • H2 and CO in UHP Ammonia matrix gas
    (see Fig. 1).

Performance*
Some standard detection limits (in parts per billion)

 

Matrix Gas

Ammonia

Impurity    
CO: Carbon Monoxide                           1
H2: Hydrogen                          0.5

All performance specifications are based on fully optimized PP1 RCP with 1 cc sample loop on continuous analysis.

 

- Unless specified, carrier gas is purified nitrogen
- Helium matrix spec is based on purified Helium as carrier
- Based on medical grade air as carrier

 

 

Accuracy

  • + / - detection limit or 10% of reading, whichever is higher.

Range

  • 1000 : 1
    Examples
    1 ppb – 1 ppm w/1m L sample loop
    10 ppb – 10 ppm w/100 ul sample loop

Operation

Run time:

  • 2 < minutes (depending on application)

Operating Temperature:

  • 55 - 85 °F (13 - 30 °C )

Carrier Gas Supply:

  • 99.999% or purified grade to < 1 ppm total impurities
  • Supply pressure 70 - 110 psig with 5% stability

Data Collection / Communication (Standard)

  • Backlit 3" x 5.25" touch screen
  • Visual chromatogram and numerical results
  • On-board integration with re-run capability
  • Chromatograms collected and archived via standard 9 – pin com-port in .csv format (data is excel compatible)

Dimensions / Electrical

  • 27" L x 17" W x 7" H
  • 25 lbs.
  • 115 VAC, 50 – 60 Hz / 220 VAC, 50-60 Hz
  • 1.5 amp maximum

Accessories and other features

  • Trend outputs
  • Analysis averaging
  • Auto start/stop capability
  • On column syringe injector adapter
  • Dual sample stream option
  • 4-20 mA Output
                  • Figure 1. (Ammonia Samples)

 

910-135

Carrier: Purified N2
Sample :Hydrogen and Carbon Monoxide within Ammonia gas samples

 

 

 

 

 

 

 

Please contact Peak Laboratories here or contact a local sales representative for custom applications and further information on our standard applications.

 