Peak Performer Gas Analysis


PEAK PERFORMER 1 PDD

Highly sensitive Pulse Discharge Detector (PDD).

The Peak Performer 1 PDD uses a highly sensitive Pulse Discharge Detector (PDD) which has been designed specifically for process applications in UHP gases.


Peak Labs Data Sheet Performer TCD

The PDD is used to measure trace impurities down to part per billion levels, while offering a common operating platform for simple, accurate measurements.

The PP1 technology has proven to offer a more cost-effective and user friendly alternative to other process GC's. In addition, Peak Labs practical experience and ability to customize to suit your applications makes Peak Labs your partner, not just your supplier.

Please contact Peak Laboratories here or contact a local sales representative for custom applications and further information on our standard applications.


Contact Us Online


Technical information

technical information tab

The Peak Performer 1 PDD gas chromatograph (GC) has been optimized for nitrogen as an impurity in UHP gases to simplify the chromatography and consistency of the process application. Typical applications are in UHP process gases such as, Ar, O2, He, and H2.


Performance*
Some standard detection limits (in parts per billion)

 

Matrix Gas

He

AR

O2

H2

N2

N20

Impurity              
N2: Nitrogen   10 10 10 10 N/A 10
Ar: Argon   5 N/A 10 10 5 N/A

All performance specifications can be optimized at best efforts for specific applications.

  • Unless specified, carrier gas is purified research grade belium

Accuracy

  • + / - detection limit or 10% of reading, whichever is higher.

Range

  • 10000 : 1
    Examples
    1 ppb – 10 ppm w/1 cc sample loop

Operation

Run time:

  • 400 seconds < (depending on application)

Operating Temperature:

  • 55 - 85 °F (13 - 30 °C )

Gas Supply Requirements:

  1. Carrier:
    • Purified research grade helium
    • Supply pressure 70 – 110 psig with 5% stability to purifier
    • Total flow ~200 sccm (plu ~ 50 cc per actuation)
  2. Actuator:
    • Purified research grade helium or equivalent regulated @ 70 psig

Data Collection / Communication (Standard)

  • Backlit 3" x 5.25" touch screen
  • Visual chromatogram and numerical results
  • On-board integration with re-run capabilitiy
  • Chromatograms collected and archived via standard 9 - pin com-port in .csv format (data is excel compatible)
  • Analog (0-1 VDC) trending available

Dimensions / Electrical

  • 27" L x 17" W x 7" H
  • 25 lbs.
  • 115 VAC, 50 – 60 Hz / 220 VAC, 50-60 Hz
  • 1.5 amp maximum

Accessories and other features

  • trend outputs
  • analysis averaging
  • auto start/stop capability



Figure 1a and Figure 1b: Example Ar / N2 analysis in Helium    



Figure 2: Step Challenge Data    


Peak Labs Datasheet

Please contact Peak Laboratories here or contact a local sales representative for custom applications and further information on our standard applications.


Contact Us Online






CLOSE WINDOW